TDS5000B Online Programmer Guide
-- 3 8 6 --
LIMit:HARDCopy
Description
This command sets or queries whether a hardcopy operation is executed
on the waveform when any waveform data exceeds the limit set in the
limit test. LIMit:STATE must be set to ON for the hardcopy operation
to execute. The hardcopy operation uses the port and prints in the for-
mat and layout specified by the HARDCopy commands.
Group
Limit Test
Related Commands
HARDCopy:FILEName, HARDCopy:PORT, LIMit:STATE
Syntax 1
LIMit:HARDCopy {OFF|ON|<NR1>}
Syntax 2
LIMit:HARDCopy?
Arguments
H OFF
This argument disables the hardcopy operation.
H ON
This argument turns on the hardcopy operation for the waveform
when any waveform data exceeds the limits set by the limit test.
H NR1
A 0 disables the hardcopy operation; any other value turns on the
hardcopy operation for the waveform when any waveform data ex-
ceeds the limits set by the limit test. For queries, a 0 is returned if the
hardcopy operation is off; a 1 is returned if the hardcopy operation is
on.
Example 1
LIMit:HARDCopy ON
This command specifies that the hardcopy operation will occur for the
waveform when any waveform data exceeds the limits specified in the
limit test.
Example 2
LIMit:HARDCopy?
This query might return:
:LIMIT:HARDCOPY 0
This indicates that the harcopy operation is disabled.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals