TDS5000B Online Programmer Guide
-- 7 7 1 --
TRIGger:A:VIDeo:SCAN
Description
This command sets or queries the video trigger horizontal line scan
rate. This command is for compatibility with earlier instruments in the
TDS series. This command is the same as the TRIGger:A:VIDeo:CUS-
Tom:SCAN command.
Group
Trigger
Related Commands
Syntax 1
TRIGger:A:VIDeo:SCAN
{RATE1|RATE2|RATE3|RATE4|RATE5}
Syntax 2
TRIGger:A:VIDeo:SCAN?
Arguments
H RATE1
This argument sets the range of the video line scan rate to 15 kHz
through 20 kHz. This is the standard broadcast rate.
H RATE2
This argument sets the range of the video line scan rate to 20 kHz
through 25 kHz.
H RATE3
This argument sets the range of the video line scan rate to 25 kHz
through 35 kHz.
H RATE4
This argument sets the range of the video line scan rate to 35 kHz
through 50 kHz.
H RATE5
This argument sets the range of the video line scan rate to 50 kHz
through 65 kHz.
Example 1
TRIGger:A:VIDeo:SCAN?
This query might return
:TRIGger:A:VIDeo:SCAN RATE2
indicating that the video line rate for the A trigger is set to Rate 2,
which is 20 kHz to 23 kHz.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals