TDS5000B Online Programmer Guide
-- 5 5 0 --
MEASUrement:IMMed:SOURCE2
Description
This command sets or queries the source to measure ”to” for phase or
delay immediate measurements. This command is equivalent to select-
ing Measurement Setup from the Measure menu, choosing the Time
tab, clicking the Delay button to display the delay settings and then
clicking the desired Source2 (To) setting.
Tip: Source2 measurements only apply to phase and delay measure-
ment types, which require both a target (Source1) and reference
(Source2) source.
Group
Measurement
Related Commands
MEASUrement:IMMed:SOURCE[1]
Syntax 1
MEASUrement:IMMed:SOURCE2
{CH<x>|MATH<y>|REF<x>|HIStogram}
Syntax 2
MEASUrement:IMMed:SOURCE2?
Arguments
H CH<x>
This is an input channel waveform. The x variable can be expressed
as an integer ranging from 1 through 4 for 4--channel models and 1 to
2 for 2--channel models.
H MATH<y>
This is a math waveform. The y variable can be expressed as an inte-
ger ranging from 1 through 4 for 4--channel models and 1 to 2 for
2--channel models.
H REF<X>
This is a reference waveform. The x variable can be expressed as an
integer ranging from 1 through 4 for 4--channel models and 1 to 2 for
2--channel models.
H HIStogram
This indicates histogram as the object to be measured.
Example 1
MEASUrement:IMMed:SOURCE2 REF3
This command sets the waveform in reference memory location 3 as the
delay ”to” source when making delay measurements.