TDS5000B Online Programmer Guide
-- 6 0 9 --
MEASUrement:REFLevel:PERCent:MID[1]
Description
This command sets or queries the percent (where 100% is equal to
HIGH) that is used to calculate the mid reference level when MEA-
SUrement:REFLevel:METHod is set to Percent. This command affects
the results of period, frequency, delay, and all cyclic measurements and
is equivalent to selecting Reference Levels from the Measure menu
and, then entering the Percentage Mid Ref value.
Note: this command affects the associated reference level parameter
for all MEASurements:IMMed and the eight periodic measurements.
To change the parameter for individual measurements, use the MEA-
SUrement:MEAS<x>:REFLevel commands.
Group
Measurement
Related Commands
MEASUrement:REFLevel:METHod
Syntax 1
MEASUrement:REFLevel:PERCent:MID[1] <NR3>
Syntax 2
MEASUrement:REFLevel:PERCent:MID[1]?
Argument
H <NR3>
This is the mid reference level, ranging from 0 to 100%. The default
mid reference level is 50%.
Example 1
MEASUrement:REFLevel:PERCent:MID[1] 60
This command sets the mid reference level to 60% of HIGH.
Example 2
MEASUrement:REFLevel:PERCent:MID[1]?
This query might return
:MEASUREMENT:REFLEVEL:PERCENT:MID 65
indicating that the percentage mid reference level is set to 65% of
HIGH.