TDS5000B Online Programmer Guide
-- 1 0 7 --
CALibrate:INTERNal:STARt
Description
This command (no query form) starts the internal signal path calibra-
tion (SPC) of the instrument. This command is the same as the CALi-
brate:INTERNal command. You can use the CALibrate:INTER-
Nal:STATus? query to return the current status of the internal signal
path calibration of the instrument.
Group
Calibration
Related Commands
CALibrate:RESults:SPC?
Syntax
CALibrate:INTERNal:STARt
Example
CALibrate:INTERNal:STARt
This command starts the internal signal path calibration of the instru-
ment.