TDS5000B Online Programmer Guide
-- 7 2 0 --
TRIGger:A:PULse:RUNT:THReshold:BOTh
Description
This command (no query form) sets the upper and lower switching
thresholds for the A pulse runt trigger. This command is equivalent to
selecting Runt Setup from the Trig menu and then setting the Upper
Level and Lower Level voltages.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:THReshold?
Syntax
TRIGger:A:PULse:RUNT:THReshold:BOTh {TTL|ECL}
Arguments
H TTL
This sets the upper threshold to 1.8 V and the lower threshold to 800
mV, which are the nominal TTL voltage levels.
H ECL
This sets the upper threshold to --1.1 V and the lower threshold to --1.5
V, which are the nominal ECL voltage levels.
Example
TRIGger:A:PULse:RUNT:THReshold:BOTh TTL
This command sets the threshold of the pulse runt trigger to the nomi-
nal TTL voltage levels.