TDS5000B Online Programmer Guide
-- 2 9 0 --
FACtory
Description
This command (no query form) resets the instrument to its factory de-
fault settings.
This command is equivalent to pressing the DEFAULT SETUP button
located on the instrument front panel or selecting Recall Default Setup
from the File menu.
Group
Save and Recall
Related Commands
*PSC, *RCL?, RECall:SETUp, *RST, *SAV, SAVe:SETUp
Syntax
FACtory
Arguments
None
Example
FACtory
This command resets the instrument settings to factory defaults:
H Clears the Event Status Enable Register.
H Clears the Service Request Enable Register.
H Sets the Device Event Status Enable Register to 255.
H Sets the Power On Status Clear Flag to TRUE.
H Purges all defined aliases.
H Enables all Command Headers.
H Sets the macro defined by *DDT to a ”zero--length field.”
H Clears the pending operation flag and associated operations.
This command does not reset the following:
H The state of the GPIB (IEEE Std 488.1--1987) interface.
H The selected GPIB address.
H Calibration data that affects device specifications.
H Protected user data.
H Stored settings.
H The current password (if you set one).

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals