TDS5000B Online Programmer Guide
-- 6 7 5 --
TRIGger:A:HOLDoff:TIMe
Description
This command sets or queries the A trigger holdoff time. This com-
mand is equivalent to selecting Holdoff from the Trig menu and then
choosing the desired Trig Holdoff.
Group
Trigger
Related Commands
TRIGger:A:HOLDoff:BY
Syntax 1
TRIGger:A:HOLDoff:TIMe <NR3>
Syntax 2
TRIGger:A:HOLDoff:TIMe?
Argument
H <NR3>
This specifies the holdoff time in seconds. The range is from 250 ns
through 12.0 s.
Example 1
TRIGger:A:HOLDoff:TIME?
This query might return
:TRIGGER:A:HOLDOFFTIME 1.2000E-06
indicating that the A trigger holdoff time is set to 1.2 µs.
Example 2
TRIGger:A:HOLDoff:TIMe 10
This command sets the A trigger holdoff time to 10 s.