-- 3 9 0 --
LIMit:SAVEWFM
Description
This command sets or queries whether the source waveform is saved
when the source waveform data exceeds the test limits.
Group
Limit Test
Related Commands
Syntax 1
LIMit:SAVEWFM {OFF|ON|NR1}
Syntax 2
LIMit:SAVEWFM?
Arguments
H OFF
This argument disables saving the source waveform when it exceeds
the test limits.
H ON
This argument enables saving the source waveform when it exceeds
the test limits.
H NR1
A 0 disables the save waveform feature so that when the source wave-
form data exceeds the limits set by the limit test, the source waveform
is not saved; any other value enables the save waveform feature so
that when the source waveform data exceeds the limits set by the limit
test, the source waveform is saved. For queries, a 0 is returned if the
save waveform feature is off; a 1 is returned if the save waveform
feature is on.
Example 1
LIMit:SAVEWFM 1
This command enables the save waveform feature so that when the
source waveform data exceeds the limits set by the limit test, the source
waveform is saved.
Example 2
LIMit:SAVEWFM?
This query might return:
:LIMIT:SAVEWFM 0
This indicates that the save waveform feature is disabled and that the
source waveform will not be saved when it exceeds the limits of the
limit test.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals