TDS5000B Online Programmer Guide
-- 4 9 8 --
MASK:USER:WIDth
Description
This command sets or returns the nominal bit width in seconds. The
query form of this command returns the bit width value of the displayed
mask. The set form of this command only affects the User mask, re-
gardless of the current (displayed) mask.
Group
Mask
Related Commands
MASK:USER:TRIGTOSAMP
Syntax 1
MASK:USER:WIDth <NR3>
Syntax 2
MASK:USER:WIDth?
Argument
H <NR3>
Is a floating point number that sets the nominal bit width in seconds.
This is the time of one bit of data where bit width = 1 / ( data rate of
the signal ).
Example
MASK:USER:WIDTH 648e-9
Sets the User mask nominal bit width to 648 nanoseconds (1.544 Mbits/
second).