-- 3 9 7 --
LIMit:TEMPLate:TOLerance:VERTical
Description
This command sets or queries the amount, in units of vertical divisions,
by which the source waveform is varied vertically when creating the
destination waveform.
Group
Limit Test
Related Commands
Syntax 1
LIMit:TEMPLate:TOLerance:VERTical <tolerance>
Syntax 2
LIMit:TEMPLate:TOLerance:VERTical?
Argument
H <tolerance>
This argument is the amount, in vertical divisions, by which the cur-
rent source waveform is allowed to deviate from the template wave-
form without exceeding the limits set in the limit test. The range is 0
to 5 divisions.
Example 1
LIMit:TEMPLate:TOLerance:VERTical 1.0
This command specifies that the current waveform is close enough to
the template waveform if it is within ±1.0 vertical division.
Example 2
LIMit:TEMPLate:TOLerance:VERTical?
This query might return:
:LIMit:TEMPLate:TOLerance:VERTical 2.0
This indicates that the current waveform is close enough to the template
waveform if it is within ±2.0 vertical divisions.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals