TDS5000B Online Programmer Guide
-- 3 9 7 --
LIMit:TEMPLate:TOLerance:VERTical
Description
This command sets or queries the amount, in units of vertical divisions,
by which the source waveform is varied vertically when creating the
destination waveform.
Group
Limit Test
Related Commands
Syntax 1
LIMit:TEMPLate:TOLerance:VERTical <tolerance>
Syntax 2
LIMit:TEMPLate:TOLerance:VERTical?
Argument
H <tolerance>
This argument is the amount, in vertical divisions, by which the cur-
rent source waveform is allowed to deviate from the template wave-
form without exceeding the limits set in the limit test. The range is 0
to 5 divisions.
Example 1
LIMit:TEMPLate:TOLerance:VERTical 1.0
This command specifies that the current waveform is close enough to
the template waveform if it is within ±1.0 vertical division.
Example 2
LIMit:TEMPLate:TOLerance:VERTical?
This query might return:
:LIMit:TEMPLate:TOLerance:VERTical 2.0
This indicates that the current waveform is close enough to the template
waveform if it is within ±2.0 vertical divisions.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold