EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #744 background imageLoading...
Page #744 background image
TDS5000B Online Programmer Guide
-- 7 2 4 --
TRIGger:A:PULse:RUNT:WHEn
Description
This command sets or queries the type of pulse width the trigger checks
for when it detects a runt. This is equivalent to selecting Runt Setup
from the Trig menu and choosing the desired Trigger When setting
from the drop--down list.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:WIDth
Syntax 1 (Four--channel instruments)
TRIGger:A:PULse:RUNT:WHEn {LOGIc|OCCurs|WID-
ERthan}
Syntax 2 (Two--channel instruments)
TRIGger:A:PULse:RUNT:WHEn {OCCurs|WIDERthan}
Syntax 3
TRIGger:A:PULse:RUNT:WHEn?
Arguments
H LOGIc
This argument specifies a trigger event when a runt occurs and the
AND of the logic channels. This argument is not available for two--
channel instruments.
H OCCurs
This argument specifies a trigger event if a runt of any detectable
width occurs.
H WIDERthan
This specifies a trigger event if a runt greater than the specified width
occurs.
Example 1
TRIGger:A:PULse:RUNT:WHEn?
This query might return
:TRIGGER:A:PULSE:RUNT:WHEN OCCURS
indicating that a runt trigger will occur if the instrument detects a runt
of any detectable width.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals