TDS5000B Online Programmer Guide
-- 7 2 4 --
TRIGger:A:PULse:RUNT:WHEn
Description
This command sets or queries the type of pulse width the trigger checks
for when it detects a runt. This is equivalent to selecting Runt Setup
from the Trig menu and choosing the desired Trigger When setting
from the drop--down list.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:WIDth
Syntax 1 (Four--channel instruments)
TRIGger:A:PULse:RUNT:WHEn {LOGIc|OCCurs|WID-
ERthan}
Syntax 2 (Two--channel instruments)
TRIGger:A:PULse:RUNT:WHEn {OCCurs|WIDERthan}
Syntax 3
TRIGger:A:PULse:RUNT:WHEn?
Arguments
H LOGIc
This argument specifies a trigger event when a runt occurs and the
AND of the logic channels. This argument is not available for two--
channel instruments.
H OCCurs
This argument specifies a trigger event if a runt of any detectable
width occurs.
H WIDERthan
This specifies a trigger event if a runt greater than the specified width
occurs.
Example 1
TRIGger:A:PULse:RUNT:WHEn?
This query might return
:TRIGGER:A:PULSE:RUNT:WHEN OCCURS
indicating that a runt trigger will occur if the instrument detects a runt
of any detectable width.

Table of Contents