-- 7 2 3 --
TRIGger:A:PULse:RUNT:THReshold?
Description
This query--only command returns the upper and lower thresholds for
the A pulse runt trigger. This command query is equivalent to selecting
Runt Setup from the Trig menu and then viewing the Upper Level and
Lower Level settings.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:POLarity,
TRIGger:A:PULse:RUNT:THReshold:BOTh,
TRIGger:A:PULse:RUNT:THReshold:HIGH,
TRIGger:A:PULse:RUNT:THReshold:LOW
Syntax
TRIGger:A:PULse:RUNT:THReshold?
Example
TRIGger:A:PULse:RUNT:THReshold?
This query might return
:TRIGGER:A:PULSE:THRESHOLD:HIGH 2.0000E+00;
LOW 8.0000E-01
indicating that the upper threshold is 2 V and that the lower threshold is
800 mV.