TRIGger:A:LOGIc:INPut:CH<x>
Description
This command sets or queries the A logical input for the logic trigger
channel specified by x. The value of x ranges from 1 through 3 for
four--channel instruments. For two--channel instruments x can only be
1. Note that CH4 on four--channel instruments or CH2 on two--channel
instruments cannot be set or queried with this command. For details
about setting this channel, see TRIGger:A:LOGIc:PATtern:INPut:CH4
or TRIGger:A:LOGIc:PATtern:INPut:CH2. This command is equiva-
lent to selecting A Event (Main) Trigger Setup from the Trig menu and
then choosing the desired logical input from the Ch<x> drop--down list,
which is located in the Input Threshold group box.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:PATtern:INPut:CH4, TRIGger:A:LOGIc:PAT-
tern:INPut:CH2
Syntax 1
TRIGger:A:LOGIc:INPut:CH<x> {HIGH|LOW|X}
Syntax 2
TRIGger:A:LOGIc:INPut:CH<x>?
Arguments
H HIGH
This specifies the logic high.
H LOW
This specifies the logic low.
H X
This specifies a ”do not care” state.
Example 1
TRIGger:A:LOGIc:INPut:CH1?
This query might return
:TRIGGER:LOGIC:INPUT:CH1 X
indicating that the setting for the A logic trigger input to channel 1 does
not matter.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals