TDS5000B Online Programmer Guide
-- 1 5 5 --
CURSor:VBArs:UNIts
Description
This command sets or queries the units for the vertical bar cursors.
Group
Cursor
Syntax 1
CURSor:VBArs:UNIts {SEConds|HERtz}
Syntax 2
CURSor:VBArs:UNIts?
Arguments
H SECOnds
Sets the units of the vertical bar cursors for the time domain (sec-
onds).
H HERtz
Sets the units of the vertical bar cursors for the frequency domain
(hertz).
Returns
A string for SECOnds or HERtz, depending on the current vertical bar
cursor units.
Example 1
CURSor:VBArs:UNIts HERtz
Sets the units of the VBArs cursors to 1/seconds.
Example 2
CURSor:VBArs:UNIts?
This command might return
:CURSOR:VBARS:UNITS SECONDS
indicating that the units for the vertical bar cursor are currently set to
seconds.