TDS5000B Online Programmer Guide
-- 7 7 7 --
TRIGger:B
Description
This command sets the B trigger level to 50% of minimum and maxi-
mum. The query form of this command returns the B trigger parame-
ters. This command is similar to selecting B Event (Delayed) Trigger
Setup from the Trig menu and then viewing the current setups.
Group
Trigger
Related Commands
TRIGger:A
Syntax 1
TRIGger:B SETLevel
Syntax 2
TRIGger:B?
Argument
H SETLevel
This sets the B trigger level to 50% of MIN and MAX.
Example 1
TRIGger:B?
This query might return the following B trigger parameters:
:TRIGGER:B:STATE 0;TYPE EDGE;
LEVEL -220.0000E-3;BY TIME;EDGE:SOURCE
CH1;SLOPE RISE;COUPLING DC;
:TRIGGER:B:TIME 16.0000E-9;EVENTS:COUNT 2
Example 2
TRIGger:B SETLevel
This command sets the B trigger level to 50% of MIN and MAX.