TDS5000B Online Programmer Guide
-- 7 2 2 --
TRIGger:A:PULse:RUNT:THReshold:LOW
Description
This command sets or queries the lower limit for the A pulse runt trig-
ger. This command is equivalent to selecting Runt Setup from the Trig
menu and then setting the Lower Level voltage.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:THReshold?
Syntax 1
TRIGger:A:PULse:RUNT:THReshold:LOW <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:THReshold:LOW?
Argument
H <NR3>
This specifies the threshold value in volts.
Example 1
TRIGger:A:PULse:RUNT:THReshold:LOW?
This query might return
:TRIGGER:A:PULSE:RUNT:THRESHOLD:LOW 1.2000E-01
indicating that the lower limit of the pulse runt trigger is set to 120 mV.
Example 2
TRIGger:A:PULse:RUNT:THReshold:LOW 50E-3
This command sets the lower limit of the pulse runt trigger to 50 mV.