DISplay:FILTer
Description
This command sets or queries the type of interpolation to use for the
display. Filtering only applies to normal--mode acquisition. The DIS-
play:FILTer command also provides selection for acquisition interpola-
tion type. This command is equivalent to selecting Waveform Interpola-
tion from the Display menu.
Group
Display Control
Syntax 1
DISplay:FILTer {LINEAr|SINX}
Syntax 2
DISplay:FILTer?
Arguments
H LINEAr
This specifies linear interpolation, where acquired points are con-
nected with straight lines.
H SINX
This specifies sin(x)/x interpolation, where acquired points are fit to a
curve.
Example 1
DISplay:FILTer SINX
This command specifies sine--curve interpolation, when magnifying
waveforms.
Example 2
DISplay:FILTer?
This query might return:
:DISPLAY:FILTER LINEAR
This indicates that straight--line interpolation is specified for magnify-
ing waveforms.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet