EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #564 background imageLoading...
Page #564 background image
TDS5000B Online Programmer Guide
-- 5 4 4 --
MEASUrement:IMMed:REFLevel:PERCent:MID[1]
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the mid reference level when MEASUre-
ment:IMMed:REFLevel:METHod is set to Percent. Note that this com-
mand affects the results of period, frequency, delay, and all cyclic mea-
surements.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:PERCent:MID[1]
<NR1>
Syntax 2
MEASUrement:IMMed:REFLevel:PERCent:MID[1]?
Arguments
H <NR1>
This is the mid reference level, ranging from 0 to 100%. The default
mid reference level is 50%.
Example 1
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID[1] 60
This command sets the mid reference level to 60% of HIGH.
Example 2
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID[1]?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID 65
Indicating that the percentage mid reference level is set to 65% of
HIGH.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals