TDS5000B Online Programmer Guide
-- 5 4 4 --
MEASUrement:IMMed:REFLevel:PERCent:MID[1]
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the mid reference level when MEASUre-
ment:IMMed:REFLevel:METHod is set to Percent. Note that this com-
mand affects the results of period, frequency, delay, and all cyclic mea-
surements.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:PERCent:MID[1]
<NR1>
Syntax 2
MEASUrement:IMMed:REFLevel:PERCent:MID[1]?
Arguments
H <NR1>
This is the mid reference level, ranging from 0 to 100%. The default
mid reference level is 50%.
Example 1
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID[1] 60
This command sets the mid reference level to 60% of HIGH.
Example 2
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID[1]?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID 65
Indicating that the percentage mid reference level is set to 65% of
HIGH.