TDS5000B Online Programmer Guide
-- 4 0 7 --
MASK:AUTOSet:MODe
Description
This command controls whether or not an autoset will be automatically
done after a standard mask is selected. The autoset will never happen if
the standard mask is selected from GPIB, since a ”SET?” sent back to
the scope should not do a mask autoset.
The default value is: manual.
This command, like all the MASK:AUTOSET commands, affects only
an autoset on a standard mask, not the general oscilloscope autoset
function.
Group
Mask
Syntax
MASK:AUTOSet:MODe {MANual|AUTO}
Argument
H MANual
This disables autoset from occurring automatically after a standard
mask is selected.
H AUTO
This enables autoset to automatically occur after a standard mask is
selected.
Example
MASK:AUTOSet:MODe AUTO
Enables autoset to occur automatically after a standard mask is selected.