-- 4 6 3 --
MASK:STOPOnviolation
Description
This command sets or returns the stop on violation setting. When en-
abled, this command stops waveform acquisitions on the first occur-
rence of a waveform violation. MASK:DISplay must be ON for the
oscilloscope to detect violations. In other words, it controls whether the
pass/fail test stops running when a violation is detected.
Group
Mask
Syntax 1
MASK:STOPOnviolation {OFF|ON|<NR1>}
Syntax 2
MASK:STOPOnviolation?
Arguments
H ON
Stops waveform acquisition on the first occurrence of a mask viola-
tion.
H OFF
Turns off mask test stop on violation.
H <NR1>
A 0 is an integer, which turns off stop on violation, and any other val-
ue turns on stop on violation.
Example
MASK:STOPONVIOLATION ON
Sets the oscilloscope to stop waveform acquisition on the first occur-
rence of a waveform violation.