TDS5000B Online Programmer Guide
-- 3 6 7 --
HORizontal:MAIn:SAMPLERate
Description
This command sets the horizontal sample rate to the desired number of
samples per second. The record length is automatically adjusted at the
same time to maintain a constant number of data points in each frame.
The query form of this command returns the current horizontal sample
rate. This command is equivalent to selecting Horizontal/Acquisition
Setup from the Horiz/Acq menu and then entering the desired Sample
Rate.
Group
Horizontal
Related Commands
HORizontal:RESOlution
Syntax 1
HORizontal:MAIn:SAMPLERate <NR3>
Syntax 2
HORizontal:MAIn:SAMPLERate?
Arguments
H <NR3>
This represents the size of the sample rate, which ranges from 5 S/s to
250 GS/s.
Example 1
HORizontal:MAIn:SAMPLERate 125E6
This command sets the sample rate to the sample rate increment that is
closest to 125 MS/s.
Example 2
HORizontal:MAIn:SAMPLERate?
This query might return
:HORIZONTAL:MAIN:SAMPLERATE 2.5000E+09