EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #387 background imageLoading...
Page #387 background image
TDS5000B Online Programmer Guide
-- 3 6 7 --
HORizontal:MAIn:SAMPLERate
Description
This command sets the horizontal sample rate to the desired number of
samples per second. The record length is automatically adjusted at the
same time to maintain a constant number of data points in each frame.
The query form of this command returns the current horizontal sample
rate. This command is equivalent to selecting Horizontal/Acquisition
Setup from the Horiz/Acq menu and then entering the desired Sample
Rate.
Group
Horizontal
Related Commands
HORizontal:RESOlution
Syntax 1
HORizontal:MAIn:SAMPLERate <NR3>
Syntax 2
HORizontal:MAIn:SAMPLERate?
Arguments
H <NR3>
This represents the size of the sample rate, which ranges from 5 S/s to
250 GS/s.
Example 1
HORizontal:MAIn:SAMPLERate 125E6
This command sets the sample rate to the sample rate increment that is
closest to 125 MS/s.
Example 2
HORizontal:MAIn:SAMPLERate?
This query might return
:HORIZONTAL:MAIN:SAMPLERATE 2.5000E+09
This indicates that the sample rate is currently set to 2.5 GS/s.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals