TDS5000B Online Programmer Guide
-- 711 --
TRIGger:A:PULse:GLItch:POLarity
Description
This command sets or queries the polarity for A pulse glitch trigger.
This command is equivalent to selecting Glitch Setup from the Trig
menu and then choosing the desired Polarity.
Group
Trigger
Related Commands
TRIGger:A:LEVel, TRIGger:A:PULse:GLItch:WIDth
Syntax 1
TRIGger:A:PULse:GLItch:POLarity {POSITIVe|NEG-
ative|EITher}
Syntax 2
TRIGger:A:PULse:GLItch:POLarity?
Arguments
H POSITIVe
This specifies that the instrument will only trigger when the polarity
of the glitch is positive.
H NEGative
This specifies that the instrument will only trigger when the polarity
of the glitch is negative.
H EITher
This specifies that the instrument will trigger when the polarity of the
glitch is either positive or negative.
Example 1
TRIGger:A:PULse:GLItch:POLarity?
This query might return
:TRIGGER:A:PULSE:GLITCH:POLARITY POSITIVE
indicating that the polarity of the glitch must be positive for the trigger
to occur.
Example 2
TRIGger:A:PULse:GLItch:POLarity EITHER
This command specifies that the polarity of the glitch can be either pos-
itive or negative for the trigger to occur.