TDS5000B Online Programmer Guide
-- 5 9 7 --
MEASUrement:MEAS<x>:VALue?
Description
This query--only command returns the value that has been calculated
for the measurement specified by <x>, which ranges from 1 through 8.
This command is equivalent to selecting Display Statistics from the
Measure menu and then choosing Value from the drop--down list to
display all measurement values on --screen.
Note: This is the displayed value in the onscreen display. If measure-
ment statistics are enabled, a new value is calculated with every wave-
form. In addition, this value is updated about every 1/3 second. If you
are acquiring a long acquisition record, the instrument may take lon-
ger to update.
Group
Measurement
Related Commands
MEASUrement:MEAS<x>:UNIts?, *ESR?, ALLEV?
Syntax
MEASUrement:MEAS<x>:VALue?
Example
MEASUrement:MEAS1:VALue?
This query might return
:MEASUREMENT:MEAS1:VALue 2.8740E-06
If the measurement has an error or warning associated with it, then an
item is added to the error queue. The error can be checked for with the
*ESR? and ALLEV? commands.