TDS5000B Online Programmer Guide
-- 1 9 4 --
DIAg:CONTROL:HALT
Description
This command (no query form) determines whether the next execution
of diagnostics will stop on the first diagnostic failure that occurs or will
execute the selected set of diagnostic functions. This command is
equivalent to selecting Instrument Diagnostics from the Utilities menu
andthenenablingHaltonFail.
Group
Diagnostics
Related Commands
DIAg:CONTROL:LOOP
Syntax
DIAg:CONTROL:HALT (ON|OFF|<NR1>}
Arguments
H ON
This enables the halt function, causing the execution of diagnostics to
halt at the first diagnostic failure that occurs.
H OFF
This disables the halt function, allowing the oscilloscope to execute
the entire set of diagnostics before halting, even if diagnostic failure
occurs.
H <NR1>
A 0 enables the halt function; any other value disables the halt func-
tion.
Example
DIAg:CONTROL:HALT ON
This command enables the halt function, causing the execution of diag-