TDS5000B Online Programmer Guide
-- 1 3 3 --
CH<x>:PROBEFunc:EXTDBatten
Description
Use this command to specify the input --output ratio (expressed in deci-
bel units) of external attenuation or gain between the signal and the
instrument input channels.
The channel is specified by x. The value of x can range from 1 through
4 for four--channel instruments, or 1 through 2 for two--channel instru-
ments.
The query form of this command returns the user--specified attenuation
in decibels: 1X = 0 dB, 10X = 20 dB, 100X = 40 dB, etc.
This command is equivalent to selecting Attenuation from the Vertical
menu, and then either viewing or setting Ext Att(dB).
Group
Vertical
Related Commands
CH<x>:PROBEFunc:EXTAtten
Syntax 1
CH<x>:PROBEFunc:EXTDBatten <NR3>
Syntax 2
CH<x>:PROBEFunc:EXTDBatten?
Argument
H <NR3>
This is the attenuation value, which is specified in the range from
--200.00 to 200.00 dB.
Example 1
CH3:PROBEFunc:EXTDBatten 2.5
This command specifies an external 2.5 dB attenuator on channel 3.
Example 2
CH1:PROBEFunc:EXTDBatten?
This query might return
:CH1:PROBEFUNC:EXTDBATTEEN 2.5000E+00
indicating that the attenuation for channel 1 is 2.5 dB.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals