TDS5000B Online Programmer Guide
-- 1 7 5 --
CURSor:XY:XDELta?
Description
This query--only command returns the XY cursor X delta value in rec-
tangular coordinates.
Note: This query is only valid when the CURSor:XY:READOUT is
set to RECTangular. If the readout is set to other than rectangular
coordinates, the IEEE representation of NAN (Not a Number) is re-
turned: 99.1000E+36.
Group
Cursor
Syntax
CURSor:XY:XDELta?
Example
CURSor:XY:XDELta?
This query might return
:CURSOR:XY:XDELTA 40.0000E-3
indicating that the ∆X value is 40 mV.