EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #746 background imageLoading...
Page #746 background image
TDS5000B Online Programmer Guide
-- 7 2 6 --
TRIGger:A:PULse:RUNT:WIDth
Description
This command sets or queries the minimum width for an A Pulse Runt
trigger. This command is equivalent to selecting Runt Setup from the
Trig menu and then setting the Width.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:WHEn
Syntax 1
TRIGger:A:PULse:RUNT:WIDth <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:WIDth?
Argument
H <NR3>
This specifies the minimum width, in seconds.
Example 1
TRIGger:A:PULse:RUNT:WIDth?
This query might return
:TRIGGER:A:PULSE:RUNT:WIDTH 2.0000E-09
indicating that the minimum width of a pulse runt trigger is 2 ns.
Example 2
TRIGger:A:PULse:RUNT:WIDth 15E-6
This command sets the minimum width of the pulse runt trigger to
15 µs.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals