TDS5000B Online Programmer Guide
-- 7 2 6 --
TRIGger:A:PULse:RUNT:WIDth
Description
This command sets or queries the minimum width for an A Pulse Runt
trigger. This command is equivalent to selecting Runt Setup from the
Trig menu and then setting the Width.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:WHEn
Syntax 1
TRIGger:A:PULse:RUNT:WIDth <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:WIDth?
Argument
H <NR3>
This specifies the minimum width, in seconds.
Example 1
TRIGger:A:PULse:RUNT:WIDth?
This query might return
:TRIGGER:A:PULSE:RUNT:WIDTH 2.0000E-09
indicating that the minimum width of a pulse runt trigger is 2 ns.
Example 2
TRIGger:A:PULse:RUNT:WIDth 15E-6
This command sets the minimum width of the pulse runt trigger to
15 µs.

Table of Contents