EasyManua.ls Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #711 background imageLoading...
Page #711 background image
TDS5000B Online Programmer Guide
-- 6 9 1 --
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE
Description
This command sets or queries the clock edge polarity for setup and hold
triggering. This is equivalent to selecting Setup/Hold Setup from the
Trig menu and then choosing the desired Clock Edge.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:SETHold:CLOCk:SOUrce, TRIGger:A:LOGIc:SE-
THold:CLOCk:THReshold
Syntax 1
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE {FALL|RISe}
Syntax 2
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE?
Arguments
H FALL
This specifies polarity as the clock falling edge.
H RISe
This specifies polarity as the clock rising edge.
Example 1
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE?
This query might return
:TRIGGER:A:LOGIC:SETHOLD:CLOCK:EDGE RISE
indicating that polarity is specified as the clock rising edge.
Example 2
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE RISE
This command specifies the polarity as the clock rising edge.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet

Related product manuals