-- 1 9 1 --
DELEte:SETUp
Description
This command (no query form) changes the setup to reference the fac-
tory setup instead of the specific user setup slot. The content of the set-
up slot is unchanged, but the data will no longer be accessible to you.
This command is equivalent to selecting Delete from the File menu,
and then clicking the specific setup you want to delete (user setups are
shown as Setup--User and are ordered in the list from 1 through 10, if
defined) or All Setups.
Note: The setup information cannot be recovered once it has been
deleted.
Group
Save and Recall
Related Commands
*RCL?, RECall:SETUp, *RST, *SAV, SAVe:SETUp, *SDS
Syntax
DELEte:SETUp {ALL|<NR1>}
Arguments
H ALL
This argument deletes all the stored setups.
H <NR1>
This argument specifies a setup storage location to delete. Setup stor-
age location values range from 1 through 10; using an out--of--range
value causes an error.
Example
DELEte:SETUp ALL
This command deletes all stored setups. All ten storage locations are
initialized to the factory default setup.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals