EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #402 background imageLoading...
Page #402 background image
TDS5000B Online Programmer Guide
-- 3 8 2 --
LIMit:COMpare:CH<x>
Description
This command sets or queries the template against which to compare
the waveform acquired from the channel specified by x. The template
can be a waveform saved in any of the reference locations (REF1
through REF4 for four--channel instruments or REF1 through REF2 for
two--channel instruments) or none.
Group
Limit Test
Related Commands
Syntax 1
LIMit:COMpare:CH<x> {NONe|REF<x>}
Syntax 2
LIMit:COMpare:CH<x>?
Arguments
H NONe
This argument turns off template testing for the channel specified by
CH<x>.
H REF<x>
This argument selects which reference waveform to use as the tem-
plate against which to compare the waveforms acquired by the speci-
fied channel (CH<x>).
Example 1
LIMit:COMpare:CH1 REF1
This command specifies REF1 as the template waveform against which
to compare waveforms acquired using CH1.
Example 2
LIMit:COMpare:CH2?
This query might return:
:LIMIT:COMPARE:CH2 REF4
This indicates that the waveform acquired by CH2 is compared to the
template waveform stored in REF4.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals