TDS5000B Online Programmer Guide
-- 5 8 2 --
MEASUrement:MEAS<x>:SOURCE[1]
Description
This command sets or queries the source for all single channel measure-
ments and specifies the source to measure ”from” when taking a delay
measurement or phase measurement. Measurements are specified by x,
which ranges from 1 through 8. This command is equivalent to select-
ing Measurement Setup from the Measure menu and then choosing the
desired measurement source.
Group
Measurement
Syntax 1
MEASUrement:MEAS<x>:SOURCE[1]
{CH<x>|MATH<y>|REF<x>|HIStogram}
Syntax 2
MEASUrement:MEAS<x>:SOURCE[1]?
Arguments
H CH<x>
This is an input channel waveform. The x variable can be expressed
as an integer ranging from 1 through 4 for 4--channel models and 1 to
2 for 2--channel models.
H MATH<y>
This is a math waveform. The y variable can be expressed as an inte-
ger ranging from 1 through 4 for 4--channel models and 1 to 2 for
2--channel models.
H REF<x>
This is a reference waveform. The x variable can be expressed as an
integer ranging from 1 through 4 for 4--channel models and 1 to 2 for
2--channel models.
H HIStogram
This is a histogram.
Example 1
MEASUrement:MEAS2:SOURCE[1] MATH1
This command specifies Math 1 as the measurement 2 source.