Example 1
HIStogram:BOXPcnt 30.0,25.0,70.0,75.0
This command sets the left boundary of the histogram box to 30% of
the graticule (3 divisions from the left edge); the top boundary to 25%
ofthegraticule(0.25x8=2.0divisions from the top edge); the right
boundary to 70% of the graticule (7 divisions from the left edge); and
the bottom boundary to 75% of the graticule (0.75x8=6.0divisions
from the top edge).
Example 2
HIStogram:BOXPct?
This query might return :
:HISTOGRAM:BOXPCNT
30.0000,25.1000,70.0000,75.2000.