Power analysis (option R&S RTE-K31)
R&S
®
RTE
1014User Manual 1326.1032.02 ─ 20
16.6.2 Transient response
The "Transient Response" analysis measures the response of a system to a change
from equilibrium. This response is described by different properties like the rise time,
the overshoot, the settling time, the peak time and the delay time.
Required probes:
●
One or two voltage probes
●
Current probe
16.6.2.1 Transient response results
The results of " Power Quality" measurements are provided in two ways:
●
The diagram shows the graphical presentation of:
– the voltage waveforms
– the current waveform
●
The result box displays the numeric measurement results.
To measure and display the transient response, the instrument uses the following mea-
surements and waveforms:
●
Meas 8 to measure the rise time, the overshoot and delay to trigger
●
"C 1" Cursor 3 to measure the peak time
●
"C 2" Cursor 2 to measure the delay to trigger
The used resources are listed in the "Details" tab. See also: Chapter 16.2.3, "Details
tab", on page 966.
The results describing the transient response of the system are shown in Figure 16-5
and described in Table 16-6.
Figure 16-5: Graphical presentation of the transient response properties
Output