Mask testing
R&S
®
RTE
443User Manual 1326.1032.02 ─ 20
Note: If you run the acquisition with [RUN CONT], the state of the mask test is set
to "Finished" when this number of acquisitions has been captured but the mask
testing continues until the acquisition is stopped.
10.3.5 Saving and loading masks
Mask test definitions remain on the instrument until they are changed or deleted, or
[PRESET] is performed. If you want to keep a mask test, you can save and reload
them.
To save a mask
1. Press the [MASKS] key on the front panel.
2. Select the "Test Definition" tab.
3. To save the mask file in the current directory, change the file name if needed, and
tap "Save".
You can use the automatic file name generation, see Chapter 12.3, "Autonaming",
on page 498.
4. To select the directory and enter the file name, tap "Save As".
To load a mask
1. To load the specified mask file, tap "Load."
2. To load the mask from a different file, tap "Open". Select the file from the file selec-
tion dialog box.
10.3.6 Mask testing on history acquisitions
In the same way as for running acquisitions, you can set up and perform the mask test-
ing also on history waveforms.
The requirements for mask testing on history waveforms are also the same:
●
The mask is defined, see Chapter 10.3.1.1, "Creating user masks", on page 436
and Chapter 10.3.1.2, "Modifying user masks", on page 437.
●
The mask test is defined, see Chapter 10.3.2, "Setting up a mask test",
on page 440
●
The mask display is configured, see Chapter 10.3.3, "Configuring the mask and hit
display", on page 441.
1. Perform and finish the acquisition.
2. Press [HISTORY].