Remote control commands
R&S
®
RTE
1412User Manual 1326.1032.02 ─ 20
To set the export data format, use FORMat[:DATA].
Return values:
<Data> List of values according to the format settings
Long term data:
If statistics are disabled, the current value of each long term
point is returned.
If statistics are enabled (MEASurement<m>:STATistics[:
ENABle]), seven values for each long term point are returned:
upper peak, lower peak, RMS, standard deviation, average,
number of measured results per long term point, number of
waveforms per long term point.
For measurement histograms, absolute values are returned.
See also: Chapter 12.2.5, "Result analysis", on page 493
Example:
See "Transferring long-term measurement data" on page 1059
Usage: Query only
18.16.10 Screenshots
The HCOPy subsystem and some other commands control the output of display infor-
mation for documentation purposes. The instrument allows two independent output
configurations which can be set separately with the suffix.
Note that the remote mode is intended for maximum performance. Therefore, the dis-
play does not follow the remote commmands consistantly. To get a correct screenshot,
turn the display on using SYST:DISP:UPD ON.
HCOPy:DESTination<1..2>........................................................................................... 1412
MMEMory:NAME......................................................................................................... 1413
HCOPy:DEVice<m>:LANGuage.................................................................................... 1413
HCOPy:DEVice<m>:COLor...........................................................................................1414
HCOPy:DEVice<m>:INVerse.........................................................................................1414
HCOPy:PAGE:ORIentation<1..2>.................................................................................. 1414
HCOPy:WBKG............................................................................................................ 1415
HCOPy:CMAP<m>:DEFault..........................................................................................1415
HCOPy:SSD................................................................................................................1415
HCOPy:ISBA...............................................................................................................1415
HCOPy:IMMediate<m>[:DUM].......................................................................................1416
HCOPy:IMMediate<m>:NEXT....................................................................................... 1416
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?.......................................................1416
SYSTem:COMMunicate:PRINter:ENUMerate[:NEXT]?.....................................................1417
SYSTem:COMMunicate:PRINter:SELect<1..2>............................................................... 1417
HCOPy:DESTination<1..2> <Medium>
Selects the output medium: file, printer or clipboard.
Data management