18.13.2 Waveform data
CALCulate:MATH<m>:DATA:STYPe?............................................................................ 1328
CALCulate:MATH<m>:DATA:HEADer?...........................................................................1328
CALCulate:MATH<m>:DATA[:VALues]?..........................................................................1329
CALCulate:MATH<m>:DATA:STYPe?
Returns the signal type of the source of the math waveform.
Suffix:
<m>
.
1..8
Selects the math waveform.
Return values:
<SignalType> SOUR | SPEC | CORR | MEAS | NONE
SOURce = normal signal
SPECtrum = FFT spectrum, specific math signal
CORRelation = correlated signal, specific math signal
MEAsurement = result of a measurement
NONE = undefined
Usage: Query only
Asynchronous command
CALCulate:MATH<m>:DATA:HEADer?
Returns the header of math waveform data. The header contains attributes of the
waveform.
Table 18-11: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval. For most
waveforms the result is 1, for peak detect and enve-
lope waveforms it is 2. If the number is 2, the num-
ber of returned values is twice the number of sam-
ples (record length).
1
Suffix:
<m>
.
1..8
Selects the math waveform.
Example:
CALC:MATH1:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1