Triggers
R&S
®
RTE
273User Manual 1326.1032.02 ─ 20
Reset timeout / Timeout
If timeout is enabled, the instrument waits for the "Timeout" time for the specified num-
ber of B-triggers. If no trigger occurs during that time, the sequence is restarted with
the A-trigger.
Remote command:
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle] on page 1205
TRIGger<m>:SEQuence:RESet:TIMeout:TIME on page 1206
Reset event
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the trigger conditions are fulfilled. The R-
trigger condition is configured in the same way as the A-trigger. The instrument disa-
bles settings that do not fit the previous settings in the sequence.
Remote command:
TRIGger<m>:SEQuence:RESet:EVENt on page 1205
6.8 External trigger input
Except for using analog or digital input channels as trigger source, you can also use
external signals as trigger source. The external signal is connected to the external trig-
ger input, which is equipped with the Rohde & Schwarz probe interface. Thus, you can
use all Rohde & Schwarz probes to connect the external trigger signal. The only trigger
type to trigger on external signals is the edge trigger.
1. Connect the external trigger signal to the external trigger input.
External trigger input