Power analysis (option R&S RTE-K31)
R&S
®
RTE
1003User Manual 1326.1032.02 ─ 20
16.5.3.2 Configuring switching loss
For details of the configuration settings, see Chapter 16.5.3.3, "Switching loss set-
tings", on page 1003.
1. Select "Analysis" > "Power".
2. Under "Power Path", select "Switching Loss".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
4. Deskew the probes as described in Chapter 16.1.1.1, "Auto deskew", on page 955.
5. Connect the probes to the DUT as shown in the "Channels" tab:
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8. Enable the parameters you want to measure.
9. Select an"Optimize Scaling" option.
10. Tap "Execute".
On the screen you can see the measurement waveforms of the current, the voltage
and the power. Also, the result box with numeric measurement results is shown.
For details, see Chapter 16.5.3.1, "Switching loss results", on page 1000.
16.5.3.3 Switching loss settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 16.2.1, "Channels tab", on page 965.
In the "Settings" tab, you configure the switching loss parameters and display settings.
Power path