Mask testing
R&S
®
RTE
426User Manual 1326.1032.02 ─ 20
Remote command:
MTESt:CONDition on page 1333
MTESt:TOLerance on page 1334
Save / load mask test
Provides all functions to store and recall a mask test. The mask definition, defined
actions and fail conditions are stored in an R&S RTE-specific .xml file.
"Open"
Opens a dialog box where you can select the file name.
See also: Chapter 12.7, "File selection dialog", on page 511.
"Save, Save
As"
Stores the specified file.
Remote command:
MTESt:FILE:NAME on page 1334
MTESt:FILE:SAVE on page 1335
MTESt:FILE:OPEN on page 1335
MTESt:FILE:DELete on page 1335
10.2.2 Mask definition
Access: [MASKS] > "Mask Definition"
With mask definition, you define the shape of the mask - the form and position of its
limit lines. The content of the "Mask Definition" tab depends on the selected "Definition
type": "User" or "Waveform".
The "Definition type" is a common setting on the top of the tab, see "Definition Type"
on page 425.
Below, you find the specific settings:
10.2.2.1 Mask definition: user mask
Access: [MASKS] > "Mask Definition" > "Definition type" = "User"
A user mask is defined by entering the time and voltage values for all corner points of
the mask segments. A user mask has at least one segment. Complex masks can have
up to 16 segments.
An inner segment is an area defined by three or more points. Upper and lower seg-
ments limit the signal on top and bottom of the screen. They are defined by a line, the
region above or below the line is set automatically as mask segment.
Mask test settings