Remote control commands
R&S
®
RTE
1162User Manual 1326.1032.02 ─ 20
PROBe<m>:SKEState <ProbeSkew>
If enabled, the skew of all connected active probes is measured, displayed, and used
for deskewing.
Suffix:
<m>
.
1..4
The suffix is irrelevant. The setting affects all active channels.
Parameters:
<ProbeSkew> ON | OFF
*RST: OFF
Usage: Asynchronous command
18.8.12 High definition mode
HDEFinition:STATe....................................................................................................... 1162
HDEFinition:BWIDth..................................................................................................... 1162
HDEFinition:RESolution?.............................................................................................. 1163
FORMat:BORDer......................................................................................................... 1163
HDEFinition:STATe <State>
Activates the high definition mode of the instrument.
Parameters:
<State> ON | OFF
ON: high definition mode, up to 16 bit digital resolution
OFF: normal oscilloscope mode
*RST: OFF
Usage: Asynchronous command
HDEFinition:BWIDth <Bandwidth>
Sets the filter bandwidth for the high definition mode.
Parameters:
<Bandwidth> Range: 10000 to max. 2 GHz, depending on the instru-
ment bandwidth.
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
See "Bandwidth" on page 190 for bandwidth limits.