Remote control commands
R&S
®
RTE
1140User Manual 1326.1032.02 ─ 20
TRPRobe:SETup:DEGauss
PROBe<m>:SETup:DEGauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement. The
demagnetizing process takes about one second.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
Firmware/software: FW 2.50
PROBe<m>:SETup:OFFSet:STPRobe
Saves the zero adjust value in the probe box. If you connect the probe to another
channel or to another R&S RTx oscilloscope, the value is read out again.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
PROBe<m>:SETup:OFFSet:ZADJust <ZroAdjVal>
set the waveform to zero position. It corrects the effect of a voltage offset or tempera-
ture drift. To set the value by the instrument, use PROBe<m>:SETup:OFFSet:AZERo.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ZroAdjVal> Range: -100 to 100
Increment: 0.1
*RST: 0
Default unit: %
Usage: Asynchronous command
Firmware/software: FW 2.50
18.8.7.8 Probe attributes
TRPRobe:ID:SWVersion?............................................................................................. 1141
PROBe<m>:ID:SWVersion?.......................................................................................... 1141
TRPRobe:ID:PRDate?.................................................................................................. 1141
PROBe<m>:ID:PRDate?...............................................................................................1141
Acquisition and setup