Power analysis (option R&S RTE-K31)
R&S
®
RTE
971User Manual 1326.1032.02 ─ 20
10. Tap "Execute".
On the screen you can see the measurement waveforms of the current, the voltage
and the power. Also, the result box with numeric measurement results is shown.
For details, see Chapter 16.3.1.1, "Power quality results", on page 967.
16.3.1.3  Power quality settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 16.2.1, "Channels tab", on page 965.
In the "Settings" tab, you configure the power measurement parameters and display
settings.
Input frequency
Selects the input frequency of the source signal.
Remote command: 
POWer:QUALity:FREQ on page 1951
Custom frequency
Sets the user-defined frequency if the "Input frequency" is set to "Custom".
Remote command: 
POWer:QUALity:FCUS on page 1952
Optimize scaling
Selects the scaling for the display of the results.
"Autoscale the signals"
Automatically selects the most appropriate scale for the display of the
results.
"Keep present settings"
The present display settings are not changed.
Input