Network operation and remote control
R&S
®
RTE
1044User Manual 1326.1032.02 ─ 20
17.5.2.2 Remote settings
The following settings are required for remote control of the instrument via a connected
computer.
Transfer data format
Selects the data format that is used for transmission of waveform data from the instru-
ment to the controlling computer.
Waveform data can be retrieved using the following commands:
CHANnel<m>[:WAVeform<n>]:DATA[:VALues]?
CALCulate:MATH<m>:DATA[:VALues]?
REFCurve<m>:DATA[:VALues]?
DIGital<m>:DATA[:VALues]?
The content of the data stream can be defined with [FILE] > "Save/Recall > Waveforms
> Interleaved X/Y" (or EXPort:WAVeform:INCXvalues).
"Ascii"
Data values are returned in ASCII format as a list of comma separa-
ted values in floating point format.
"FLOAT"
Binary format. The data is stored as binary data (Definite Length
Block Data according to IEEE 488.2).
"INT8"
Signed integer data with length 8 bit.
"INT16"
Signed integer data with length 16 bit. The Byte order can be set
using .
For details on the formats, refer to the description of the remote com-
mand.
Remote command:
FORMat[:DATA] on page 1082
Byte order
Sets the endianess for INT16 data:
LSB first: little endian, least significant byte first
MSB first: big endian, most significant byte first
Remote command:
FORMat:BORDer on page 1083
Bit pattern format
Sets the format for all bit pattern queries.
Remote command:
FORMat:BPATtern on page 1083
SCPI logging
If enabled, all received remote commands are written into a text file. Enter the path and
filename with extension (log, txt, or csv) in "Log file path".
SCPI emulation
Not available for R&S RTE
Remote control