Protocol analysis
R&S
®
RTE
685User Manual 1326.1032.02 ─ 20
Remote command:
BUS<m>:ARINc:MAXGap:BITS on page 1634
BUS<m>:ARINc:MAXGap:SELect on page 1633
BUS<m>:ARINc:MINGap:BITS on page 1634
BUS<m>:ARINc:MINGap:SELect on page 1634
TRIGger<m>:ARINc:MINGap:BITS on page 1638
TRIGger<m>:ARINc:MINGap:SELect on page 1638
TRIGger<m>:ARINc:MAXGap:BITS on page 1638
TRIGger<m>:ARINc:MAXGap:SELect on page 1638
13.10.3.2 Triggering on ARINC 429
Prerequisites: A bus is configured for the ARINC 429 signal to be analyzed.
1. Press the [PROTOCOL] key and select the "Trigger" tab.
2. Press the "Select" button. The "Source" button is shown.
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to ARINC 429.
5. Select the "Trigger type".
6. For more complex trigger types, enter the data pattern conditions.
For details, see Chapter 13.10.3.1, "ARINC 429 trigger settings", on page 681.
13.10.4 ARINC 429 label list
Label lists are protocol-specific. An ARINC 429 label file contains two values for each
identifier:
●
"Arinc Label": the ARINC 429 label value, that identifies the data type and the
parameters associated with it. The usual data format is octal.
●
"Symbolic label": symbolic name of the label, specifying its function.
Example: ARINC 429 PTT file
# ----------------------------------------------------------------------------
@FILE_VERSION = 1.0
@PROTOCOL_NAME = arinc429
# ----------------------------------------------------------------------------
# Labels for ARINC 429 protocol
# Column order: Arinc Label, Symbolic Label
# ----------------------------------------------------------------------------
# ----Definition----
001o, Distance to Go
002o, Time to Go
010o, Present Position - Latitude
011o, Present Position - Longitude
014o, Magnetic Heading
ARINC 429 (option R&S
RTE-K7)