Triggers
R&S
®
RTE
272User Manual 1326.1032.02 ─ 20
Couple trigger levels
Sets the trigger levels to the values of the current trigger condition. Each channel has
its own trigger level.
Only available in "A → B → R" sequences.
Example:
If the "A" tab is selected in the "Setup" tab, and the trigger level for C1 is 70 mV, the
coupling sets the trigger levels for C1 in the B- and R trigger conditions also to 70 mV.
If the B-trigger and/or R-trigger uses another source as the A-trigger, the level remains
unchanged.
Remote command:
TRIGger<m>:ECOupling on page 1204
Delay
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Remote command:
TRIGger<m>:SEQuence:DELay on page 1205
B-event count
Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trig-
ger causes the trigger event.
The waiting time for B-triggers can be restricted with a reset condition: timeout or reset
event.
Remote command:
TRIGger<m>:SEQuence:COUNt on page 1205
6.7.2 R-Trigger setup
Access: [TRIGGER] > "Setup" tab > "A → B → R" sequence > "R" subtab
The reset condition R can be a timeout or a trigger condition, or a combination of both.
Sequence