Remote control commands
R&S
®
RTE
1204User Manual 1326.1032.02 ─ 20
18.9.18 Trigger sequence
TRIGger<m>:SEQuence:MODE.................................................................................... 1204
TRIGger<m>:ECOupling...............................................................................................1204
TRIGger<m>:SEQuence:DELay.................................................................................... 1205
TRIGger<m>:SEQuence:COUNt................................................................................... 1205
TRIGger<m>:SEQuence:RESet:EVENt..........................................................................1205
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle]..........................................................1205
TRIGger<m>:SEQuence:RESet:TIMeout:TIME............................................................... 1206
TRIGger<m>:SEQuence:MODE <Type>
Selects the type of the sequence.
See also: Chapter 6.7, "Sequence", on page 270.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Type> AONLy | AB
AONLy
Triggers only on A-events. Additionally, a holdoff condition can
be set. if AONLy sequence is set, all inputs (input channels,
serial and parallel buses, digital channels etc.) can be used as
trigger source.
AB
Triggers if all conditions of A- and B-events, as well as additional
delay and waiting conditions are fulfilled. This trigger sequence
requires that analog input channels CHAN1...4 are set as trigger
sources for all events.
*RST:
AONLy
Usage: Asynchronous command
TRIGger<m>:ECOupling <TrigLevEvtCpl>
Event coupling of trigger levels: Sets the trigger levels to the values of the indicated
event. Thus, channel 1 has one trigger level for all events, channel 2 has one trigger
level and so on.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<TrigLevEvtCpl> ON | OFF
*RST: ON
Usage: Asynchronous command
Trigger