Search functions
R&S
®
RTE
468User Manual 1326.1032.02 ─ 20
Remote command:
SEARch:TRIGger:LEVel:NOISe[:STATe] on page 1383
Hysteresis mode
Defines whether values absolute or relative to the vertical scaling are used.
Remote command:
SEARch:TRIGger:LEVel:NOISe:MODE on page 1382
Relative / Absolute hysteresis
Defines a range in absolute or relative values around the search level. If the signal jit-
ters inside this range and crosses the level, no search event is detected.
Absolute hysteresis values are adapted when the relative hysteresis is changed , and
vice versa.
If you change the vertical scaling, either the relative or the absolute value is adjusted
automatically.
Remote command:
SEARch:TRIGger:LEVel:NOISe:ABSolute on page 1382
SEARch:TRIGger:LEVel:NOISe:RELative on page 1383
11.5.2 Defining noise rejection for searches
1. Press the [SEARCH] key to open the "Search" dialog box.
2. Select the "Noise reject" tab.
3. Select the tab for the search you want to configure.
4. Define the absolute or relative hysteresis. If you change one value, the other is
automatically calculated.
Noise reject