Protocol analysis
R&S
®
RTE
809User Manual 1326.1032.02 ─ 20
PHYAD/PRTAD ← MDIO Trigger Type
Sets the physical address or port address (5 bits) of the frame pattern; available only in
trigger type "Data".
Remote command:
TRIGger<m>:MDIO:PHYS on page 1780
REGAD/DEVAD ← MDIO Trigger Type
Sets the register address or device address (5 bits) of the frame pattern; available only
in trigger type "Data".
Remote command:
TRIGger<m>:MDIO:REGI on page 1780
DATA/ADDRESS ← MDIO Trigger Type
Defines the payload data pattern or address pattern (16 bits); available only in trigger
type "Data".
Remote command:
TRIGger<m>:MDIO:DATA on page 1780
13.15.3.2 Triggering on MDIO
Prerequisite: A serial bus is configured for the MDIO signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [PROTOCOL] key and select the "Trigger" tab.
2. Press the "Select" button. The "Source" button is shown.
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Tap "Serial bus" and select the serial bus that is set to MDIO.
5. Tap "Trigger Type MDIO" and select the trigger type to be used for MDIO protocol
analysis.
6. If the trigger type "Data" is selected, the frame pattern has to be specified.
For information on how to proceed with the configuration settings, see Chap-
ter 13.15.3.1, "MDIO trigger settings", on page 806.
13.15.4 MDIO label list
Label lists are protocol-specific. A label list file for MDIO contains physical addresses
and their symbolic names.
MDIO (option R&S
RTE-K55)