18.9.6 Window trigger
TRIGger<m>:LEVel<n>:WINDow:LOWer........................................................................1176
TRIGger<m>:LEVel<n>:WINDow:UPPer.........................................................................1176
TRIGger<m>:WINDow:RANGe......................................................................................1176
TRIGger<m>:WINDow:TIME......................................................................................... 1177
TRIGger<m>:WINDow:WIDTh.......................................................................................1177
TRIGger<m>:WINDow:DELTa........................................................................................1178
TRIGger<m>:LEVel<n>:WINDow:LOWer <Level>
TRIGger<m>:LEVel<n>:WINDow:UPPer <Level>
Set the lower and upper voltage limits for the window.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...31 = not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
TRIGger<m>:WINDow:RANGe <RangeMode>
Defines the signal run in relation to the window:
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<RangeMode> ENTer | EXIT | WITHin | OUTSide
ENTer
Triggers when the signal crosses the upper or lower level and
thus enters the window made up of these two levels.
EXIT
Triggers when the signal leaves the window.
WITHin
Triggers if the signal stays between the upper and lower level for
a specified time. The time is defined using the TRIGger<m>:
WINDow:TIME command.
Trigger

Table of Contents